How to Plan Calibration for NABL/ISO/IEC 17025:2017
Defining the scope of work in terms of facility to be developed for the calibration of which disciplines as well as type of equipment. In this stage, the detail discussion is done with the customer and list of equipment to be calibrated is prepared with the details of name, range, resolution / least count etc.
All automated Instruments such as cell
counters, Clinical Biochemistry automated analysers, automated coagulometers
and ELISA readers shall be calibrated by manufacturer at least once a year and be
verified for calibration after preventive maintenance.
fluorimeter / LED that may / may not be dye
based, filter or emission light source lamps, pipettor assembly inclusive of
metering pump and syringe, pressure checks wherever applicable, probe alignment
and their carry over checks, temperature of temperature-controlled chambers,
cuvette calibration (wherever applicable) and system checks should be carried
out. In addition to the verification, analyte calibration may be performed and
appended.
Certain Instrument may be calibrated by the
laboratory itself without the service of an external calibration body, provided
the laboratory has necessary reference standards and materials (with
traceability) and such calibration procedures do not demand special techniques
which are outside the capability and experience of the laboratory staff. The calibration
status, date of calibration and due date shall be exhibited on the equipment.
pH meter shall be calibrated each time
before use. This is done by using two standard buffer
solutions appropriate to the expected pH of
the sample being tested and the records of the
same to be maintained.
ISE analyzers
& ABG analyzers:
Calibration verification is in the form of
self calibration. However, the electrode slope should be
recorded to ensure that there is no
deviation. Thereby requirement for change of electrodes is
not missed.
Spectrophotometer and Colorimeter:
Calibration checks on all spectrophotometers
or colorimeters shall be performed at six months /one year interval preferably
by manufacturer or AMC agency.
Electrophoresis Apparatus:
Instrument performance shall be routinely
monitored during use with appropriate controls.
System components (e.g. electrodes, tank and power supply), must be
checked periodically.
Temperature-Controlled
Equipment:
Equipment such as water baths, incubators,
ovens, refrigerators and deep freezers are verified
for accuracy / performance (with calibrated
temperature-recording devices) for the intended
temperature required. The temperature shall
be monitored daily, especially where the
deterioration of material (sample or reagents) affects the test results.
Comments
Post a Comment